Pular para o conteúdo principal
GE Healthcare · CTLightSpeed1831839608221092218318465001183100

LightSpeed 3.X (Ultra, Plus, QX/i) - Octane

Symptom

Scanner Hardware Stopped Scan message during axial scan. Long Description for Feedback Only. Scanner Hardware Stopped Scan message during axial scan

Error message

Error: 183183 kV out of 3 % tol See Long Description for error example Error: 183183 Suite: CT01 Host: OBC Proc: Generator File: gen tech.c 1.32 Method: No Method Line: 791 Function: XRAY GENERATION : KV Loop 960822 1230 10922 Scan: 0/0/3 Type: Static Exception Level: Sec/Soft Time: 12:17:46:610 Log Series: 529 kV out of 3 % tol Selected technic: 100 kV, 50 mA Tot. samples = 3 Tot. kV samples above tol. = 0, below tol. = 1 Avg. of kV samples above tol. = 0.000 kV, below tol. = 53.871 kV 960822 1230 10922 Avg. of tot. kV = 84.396 kV

Error: 183184 mA out of 5 % tol Open Long Description for full error Tue Jun 7 03:35:46 2005 Error: 183184 Suite: GEC2 Host: OBC Process: Generator File: gen_tech.c 1.36 Method: No Method Line: 924 Function: XRAY GENERATION : MA Loop Scan: 65001/11/14 Type: Axial Exception Level: Sec/Soft Time: 03:35:45:600 Log Series: 223 mA out of 5 % or 2 mA tol Selected technic: 140 kV, 56 mA Tot. samples = 4 Tot. mA samples above tol. = 0, below tol. = 1 Avg. of mA samples above tol. = 0.0 mA, below tol. = 0.2 mA Avg. of tot. mA = 26.7 mA

Error: 183100 Two consecutive kV samples out 10.0 % tolerance Open Long Description to see error example Tue Jun 7 03:35:46 200

Applies to

Models
LightSpeed
Product
LightSpeed 3.X (Ultra, Plus, QX/i) - Octane Always 4 or 8 slice MDAS 308i.2_H3.1M5 LightSpeed 3.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 308I.2_H3.1M5 305.4_H3.1M4 - obsolete 303.1_H3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
46-170021P92 SEMICONDUCTOR FUSE - 60A 700V Part #: 46-170021P92 Part Type: GP CONSUMABLES Part Weight: 0.45 LBR Repairable: No Returnable: No Hazardous Material: N Description: SEMICONDUCTOR FUSE FRU: Yes Status: Orderable Comments: FLS - CLASS 1 FRU INACTIVE TO MFG. CJB 1/94 60A 0300V LIMTD RANGE SLO W UL/C DUAL ELEMENT. (REVIEWED TK, 7/91 - OK FUS RAP 60A 700V FBP60 FLS - FE ASSIST: IN ORDER TO INCREAS INCREASE REVENUE GLOBAL PARTS HAS CREATED A PROGRAM TO HARVEST PARTS FROM RENEWABLE RESOURCRESOURCES. THESE PARTS OFFER DIRECT SAVINGS TO THE BUSINESSAND THE FIELD. ORDER THE U PART WHEN AVAILABLE. IF NO U USED PARTS ARE AVAILABLE CARES WILL CHANGE THE ORDER TOA NEW PART. SEMICONDUCTOR FUSE FLS - CLASS 1 FRU INACTIVE TO MFG. CJB 1/94 60A 0300V LIMTD RANGE SLO W UL/C DUAL ELEMENT. (REVIEWED TK, 7/91 - OK FUS RAP 60A 700V FBP60
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required