Unable to Scan with Inductive or Direct Drive Bias Fault Due to DD bias cable was disconnected from J74 on the magnet ro…
Unable to Scan with Inductive or Direct Drive Bias Fault Due to DD bias cable was disconnected from J74 on the magnet room side of the penetration panel.
Symptom
Unable to Scan with Inductive or Direct Drive Bias Fault Unable to Scan with Inductive or Direct Drive Bias Fault. Sat Apr 23 10:21:22 2005 Error: 2225526 Host: lx-mr Process: NSP File: HPC_PORT_ID:spi_tr_mon Line: 6100 Inductive or Direct Drive fault, or Body TR switch fault due to impedance mismatch, path short Drive bias fault. Possibly a shorted diode in the driver switches, or a short circuit load.
Error message
2225525 Inductive or Direct Drive fault. Driver bias fault Thu Feb 6 04:02:38 1997 2225525 1 lx-bay14b NSP HPC_PORT_ID:spi_tr_mon 6090 Inductive or Direct Drive fault. Driver bias fault. Possibly a disconnected cable between the driver board and the driver switches, or an open diode in the driver switches.
Applies to
- Models
- Signa
- Product
- Signa LX Signa LX Product can be used with 8.x and 9.x systems.
- Software
- 9.1 For Signa LX product only
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.
- CWA ID100315. Auto and Manual Prescan Fails.
- What is the Acoustic Noise Level in the Ovation Magnet.