Pular para o conteúdo principal
GE Healthcare · MRSigna2236698

MRIQ Ghosting on FIESTA images.

Symptom

MRIQ Ghosting on FIESTA images Images were noisy and the ghosting would walk across the image

Error message

2236698 CERD Diags Analog Receiver Channel Present LO Test Failure. Receiver channel 1 is missing LO1 or LO2 per RF Receiver Status register. Error: 2236698 File: diagsMain.m Line: 2039 CERD Diags : Analog Receiver Channel Present LO Test Failure Receiver channel 1 is missing LO1 or LO2 per RF Receiver Status register. Check receiver module installation. Error: 2236698 File: diagsMain.m Line: 2039 CERD Diags : Analog Receiver Channel Present LO Test Failure. Receiver channel 1 is missing LO1 or LO2 per RF Receiver Status register. Check receiver module installation.

Applies to

Models
Signa
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only
Area
MR

Full cause & step-by-step resolution