GE Healthcare · MRSigna2238439
Will Not Scan with Gradient Errors.
Symptom
Will Not Scan with Gradient Errors.
Error message
2238439 The MIF is reporting a transformer over temperature fault on the % axis Mon Oct 8 06:44:32 2007 Error: 2238439 Host: MR1 Process: NSP ile: HPC PORT ID:GAP/GIP/GP Line: 1028 The MIF is reporting a transformer over temperature fault on the Z axis HIGH 8645 Power Module. Wait for the temperature to cool before trying to scan. Cooling time may be in excess of 10 minutes.
Applies to
- Models
- Signa
- Product
- Signa LX Signa LX Product can be used with 8.x and 9.x systems.
- Software
- 9.1 For Signa LX product only
- Component
- 46-328206P115 CROWN A10999-G2 8645 PWR MOD W/CROWB
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.