EPT B0 Dither And B0 Image Won't Run.

Symptom

B0 Dither And B0 Image in EPT Fail To Complete See long description. After performing a restoreInfo, EPT Bo Dither and B0 Image tests will give an error: "EPT was unable to determine a PASS or FAIL status of the test". Both the "B0 Dither and the B0 Image" tests attempt to open the file /usr/g/caldir/ecmt.config and read the "warp_period" parameter. If the ecmt.config file does not exist, the tool tries to open /usr/g/caldir/grafidy.config to read the warp_period parameter.

Error message

EPT was Unable to Determine pass/fail Status of the Test

Applies to

Models
Signa
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only
Area
MR