GE Healthcare · MRSigna2225754
Analogic RF Amplifier heat sink is overheated due to power supply FETS are shorted.
Symptom
Scan stops with Analogic RF Amplifier 2 Power Supply Fault error 2225754 2225754 - Analogic RF Amplifier 2 Power Supply Fault. Analogic Amplifier 2 indicates the main power supply is out of specification. Possible causes: Supply feedback loop exceeds 2.5v, PS heat sink is overheated, Power supply FETS are shorted.->
Error message
2225754 Analogic RF Amplifier 2 Power Supply Fault see long description Tue Sep 11 13:22:14 2007 Error: 2225754 Host: apm1 Process: NSP File: HPC_PORT_ID:spi_rf_mon Line: 3066 Analogic RF Amplifier 2 Power Supply Fault. Analogic Amplifier 2 indicates the main power supply is out of specification. Possible causes: Supply feedback loop exceeds 2.5v, PS heat sink is overheated, Power supply FETS are shorted.->
Applies to
- Models
- Signa
- Product
- Signa LX Signa LX Product can be used with 8.x and 9.x systems.
- Software
- 9.1 For Signa LX product only
- Component
- 2224562-2 1.5T RF AMP POWER MODULE W/ FRU PKG (SRFD)
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.
- CWA ID100315. Auto and Manual Prescan Fails.
- What is the Acoustic Noise Level in the Ovation Magnet.