Analogic RF Amplifier heat sink is overheated due to power supply FETS are shorted.

Symptom

Scan stops with Analogic RF Amplifier 2 Power Supply Fault error 2225754 2225754 - Analogic RF Amplifier 2 Power Supply Fault. Analogic Amplifier 2 indicates the main power supply is out of specification. Possible causes: Supply feedback loop exceeds 2.5v, PS heat sink is overheated, Power supply FETS are shorted.->

Error message

2225754 Analogic RF Amplifier 2 Power Supply Fault see long description Tue Sep 11 13:22:14 2007 Error: 2225754 Host: apm1 Process: NSP File: HPC_PORT_ID:spi_rf_mon Line: 3066 Analogic RF Amplifier 2 Power Supply Fault. Analogic Amplifier 2 indicates the main power supply is out of specification. Possible causes: Supply feedback loop exceeds 2.5v, PS heat sink is overheated, Power supply FETS are shorted.->

Applies to

Models
Signa
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only
Component
2224562-2 1.5T RF AMP POWER MODULE W/ FRU PKG (SRFD)
Area
MR