GE Healthcare · CTLightSpeed26000741416790642052189179672126000738650056
LightSpeed 4.X - GRE
Symptom
Scanner Hardware Stopped Scan with DAS parity errors
Error message
Error: 260007414 Questionable DAS Channel Detected Mon Aug 15 07:34:52 2005 Error: 260007414 Host: DAS Process: DCB Firmware File: dcvtst.c Line: 202 Function: DATA ACQUISITION : Scan Data Acquisition Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Soft Time: 07:34:52:365 Log Series: 12 Questionable DAS Channel Detected. DAS Output View Data did NOT Pass Selftest Converter Board Card: 44 Channel: 66. Channel Value Was: -32 Counts Expected: 1679064 to 2052189 Counts Average of All Channels Is: 1796721 Error: 260007386 DAS Has Detected An Error. Data Parity Error detected in the xx Data Stream(s) on Converter Board xx or the DCB Board. See LD for sample errors Wed Mar 23 15:46:11 2005 Error: 260007386 Host: DAS Process: DCB Firmware File: parity.c Line: 219 Function: DATA ACQUISITION : Scan Data Acquisition Scan: 214/5/1 Type: Helical Exception Level: Sec/Soft Time: 15:46:11:545 Log Series: 660 SCAN ABORT -- DAS Has Detected An Error. Data Parity Error detected in the 1B Data Stream (s) on Converter Board 33 or on Converter Board 41 . NOTE: Before troubleshooting, go through gesys.log (viewlog) and search for all “Data Parity Error” reports in the log. There are 2 items you need
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 4.X - GRE See Long Description for detailed information - H2_P_M16 Product Name: LightSpeed 4.X with GRE Product Photographs GE HEALTHCARE Method of remote or software identification - See attachment below for a complete list of software commands and typical results 1. check_config 2 . swhwinfo “swhwinfo” will have the following switches: · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console types: GOC3 (GRE) GOC4 (GRE) List of software the product can run: 07MW11.10 SP2.0, SP3.0, SP3.1 FAQ's Table weight limit - 180 kg (400 lbs) normal operation - 205 kg (450 lbs) maximum allowed with decreased accuracy Tube type - D3186T D3186T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2371627-4 replaces 2371627-3 GDAS / MDAS 16 Slice RIGHT Backplane Chassis Asm EDITION 1-2 See LD for FEMC comments 2371627-4 replaces 2371627-3 GDAS / MDAS 16 Slice RIGHT Backplane Chassis Asm EDITION 1-2 Part #: 2371627-4 Part Type: GP CONSUMABLES Part Weight: 2.18 LBR Repairable: No Returnable: No Hazardous Material: N Description: GDAS-MDAS 16 Slice RIGHT BP-Chassis Asm EDITION 1-2 FRU: Yes Status: Orderable Comments: Added as replacement for 2371627, -2, -3 & 2332630 per ecr 2012678, ms13452 050518 Part #: 2371627-3 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: BP-Chassis Right EMC Asm - GDAS FRU: Yes Status: Orderable Comments: 2371627-3 is backward compatible replacement for 2371627-2. DO NOT USE 2371627-2 to replace 2371627-3. 2371627-3 is EMC2 Compliant. ms13452 PART ADDED PER VCT RSPL AND IS EMC2 COMPLETE MS8336 12/08/2004
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.