GE Healthcare · CTLightSpeedVCT26000102234949
LightSpeed 4.X - GRE
Symptom
Unable to scan with a message on screen that says: Collimator/DAS reset failed.
Error message
Error: 260001022 Communications between the CCB sub-system and OBC have been lost. Attempting to establish communications See long description for full error Fri Feb 10 14:15:14 2006 Error: 260001022 Host: OBC Process: HOP File: hop_init_state.c Line: 249 Function: No System Function Reported Scan: 34949/2/0 Type: None/Unknown Exception Level: Pri/Most Severe Time: 14:15:13:720 Log Series: 600 Communications between the CCB sub-system and OBC have been lost. Attempting to establish communications.
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 4.X - GRE See Long Description for detailed information - H2_P_M16 Product Name: LightSpeed 4.X with GRE Product Photographs GE HEALTHCARE Method of remote or software identification - See attachment below for a complete list of software commands and typical results 1. check_config 2 . swhwinfo “swhwinfo” will have the following switches: · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console types: GOC3 (GRE) GOC4 (GRE) List of software the product can run: 07MW11.10 SP2.0, SP3.0, SP3.1 FAQ's Table weight limit - 180 kg (400 lbs) normal operation - 205 kg (450 lbs) maximum allowed with decreased accuracy Tube type - D3186T D3186T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required