Will Not Scan With Patient Comfort Sensor Trip.

Symptom

Will Not Scan With Patient Comfort Sensor Trip. See Long Description Patient comfort sensor trip; cannot scan due to bore sensor indicating temperature is too hot for patient safety. A message "Patient Comfort Sensor Trip" appears in the message box on the left side of the scan page at the start of autoprescan.

Error message

2225687 Hardware Failure: The Bore Temperature Reading is invalid. It is Reading too High. Wed Oct 9 14:49:34 2002 Error: 2225687 Host: MCW-MR2 Process: NSP File: HPC_PORT_ID:spi_bore_wall_temp_mon Line: 400 Hardware Failure: the bore temperature reading is invalid. It is reading too high. Check for an open in the sensor, sensor cable, or connector, or a problem with the amplifier board on the SRI.

Applies to

Models
Signa
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only
Area
MR