GE Healthcare · MRSigna2218079
Unable to Start Image Database Due to Software Corruption Caused by PCI Cardcage.
Symptom
Unable to scan with message Prescription allocation request fail System is able to prescan. But when the prescan ends and the scan is about to begin, it fails with a message "Prescription allocation request failed". This message doesn't appear in the log, but another one does (see below): Error: 2218079 Host: MR6 Process: msgView File: ImageStat.cc Line: 68 Image space query failed
Error message
PCI Bridge Errors. device reported parity error. PCI bridge detected parity error. PCI Bridge Error: int.status is 0x5000 PCI device reported parity error PCI bridge detected parity error Prescription Allocation Request Fail 2218079 Image space query failed
Applies to
- Models
- Signa
- Product
- Signa LX Signa LX Product can be used with 8.x and 9.x systems.
- Software
- 9.1 For Signa LX product only
- Component
- 2169940-53 PCI Card Cage For The Octane Computer
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.
- CWA ID100315. Auto and Manual Prescan Fails.
- What is the Acoustic Noise Level in the Ovation Magnet.