Pular para o conteúdo principal
GE Healthcare · MRSigna2225554

Cannot Scan With intermittent TR and Dynamic Disable Faults.

Defective coil.

Symptom

Cannot Scan With intermittent TR and Dynamic Disable Faults. The customer when scanning is getting intermittent TR and Dynamic Disable faults. It happens with all coils and cannot be pinned down to a specific PSD.

Error message

2225554 Multi-Coil T/R Driver Board Fault. An Open Circuit was Detected on Port 4. Thu Aug 21 12:06:10 1997 Suite: MRS1 Host: OC0 Proc: HPC_PORT_ID Error: 2225554 File: PMH.m 1.29 Method: spi_tr_mon Line: 6260 Multi-Coil T/R Driver Board Fault. An open circuit was detected on port 4. Check cable to multi-coil board or check for an open pin diode.

Applies to

Models
Signa
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only
Component
2215261 1.5T Torso Phased Array Flex Coil See long description for Coil Shipping list and part numbers. This information is Property of GE Healthcare - For GE Healthcare Service Personnel Only. Disclosure to Third Parties Prohibited. From the SIGNA® TORSO PHASED ARRAY COIL REV 2 DIRECTION 2218805 Documentation. To view the most recent appropriate and detailed documentation, click to go to SERVICE METHODS . 1.5T TORSO PHASED ARRAY COIL SHIPPING LIST
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare MR issues