GE Healthcare · MRSigna22279721182922279732224360
The BIT 3 FIBER OPTIC INTERFACE Module in the Octane Computer Cabinet Failed.
Symptom
Some or All Images in the Last Scan Did Not Reconstruct Some or All Images in the Last Scan Did Not Reconstruct.
Error message
2227972 Wait time to receive response from TIR exceeded 20secs...still waiting 8097 Send failed Tue Mar 29 12:11:27 2005 Error: 8097 Host: GE_MRI Process: aqserver File: aqscommand.c Line: 759 Send failed: 11829 (ADW20) 2227973 Received image transfer response from TIR 2227973 Host: lx-mr Process: NSP File: TPS:ed_wait_on_cm Line: 11 Received image transfer response from TIR 8053 Remote Socket I/O Error: Connection timed out Tue Mar 29 12:11:27 2005 Error: 8053 Host: GE_MRI Process: aqserver File: session.c Line: 174 Remote Socket I/O Error: : Connection timed out Tue Mar 29 12:11:27 2005 Error: 8053 Host: GE_MRI Process: aqserver File: session.c Line: 174 Remote Socket I/O Error: :<establish_conn> Connection timed out 2224360 Unable to communicate to the Tardis chassis after multiple attempts. Image 7208/1/2 was not installed in the database. Fri Oct 8 07:53:33 2004 Error: 2224360 Host: mr02 Process: TIR re_pixel_header.c Line: 320 Host: mr02 Process: TIR File: store_pixel_header.c Line: 320 Unable to communicate to the Tardis chassis after multiple attempts. Image 7208/1/2 was not installed in the database. Check the Bit-3 connection to the Tardis chassis if further images ar
Applies to
- Models
- Signa
- Product
- Signa LX Signa LX Product can be used with 8.x and 9.x systems.
- Software
- 9.1 For Signa LX product only
- Component
- 2149852-3 BIT3 module Fiber optic convertor used on Indigo and early Octane LX systems. Fiber optic convertor used on Indigo and early Octane LX systems.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.
- CWA ID100315. Auto and Manual Prescan Fails.
- What is the Acoustic Noise Level in the Ovation Magnet.