LightSpeed 7.x (VCT, VCT Select, Pro32)

Symptom

"Scanner Hardware Stopped Scan" with CAM or Filter errors

Error message

Error: 260133163 CAM (A or B) Move Is Out of The Retry Tolerance. See LD for complete error Sun Sep 10 15:17:52 2006 Error: 260133163 Host : Collimator Ermes # : 260133163 Exception Class : Secondary Severity : Pri/Soft File : ccb_drive_move.cxx Line# : 341 Function : No System Function Reported Scan Type : None/Unknown Scan: 5220/1/0 CAM A Move Is Out of The Retry Tolerance. Actual Position : 6115um 0cnt Expected Position: 7385um -500cnt Retry Tolerance : 150um

Error: 260133167 CAM (A or B) drive monitoring detected no motion when motion was requested. see long description for complete error code: Thu Jan 19 07:51:06 2006 Error: 260133167 Host: orp Process: orp_subsys File: ccb_drive_timers.cxx Line: 543 Thu Jan 19 07:51:06 2006 Host : Collimator Ermes # : 260133167 Exception Class : Abort Severity : Secondary File : ccb_drive_timers.cxx Line# : 543 Function : No System Function Reported Scan Type : Static Scan: 50013/1/1 CAM A drive monitoring detected no motion when motion was requested. Verify operation of stepper motor and encoder feedback Start Position = 5970um 0cnt Desired Position = 102um 1659cnt Current Motor Position = 3383um 731cnt Current Encoder Position= 5970um 0cnt

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
Component
5129998-2 ccb chassis assembly VCT-GAN-COLL This is actually the Collimator Control Board and the metal bracket only - it is NOT the complete collimator assembly. See LD for details 5129998-2 ccb chassis assembly VCT-GAN-COLL This is actually the Collimator Control Board and the metal bracket only - it is NOT the complete collimator assembly Part #: 5129998-2 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: CCB Chassis Asm, VCT Failsafe VCT-GAN-COLL FRU: Yes Status: Orderable Comments: Added by ECR 2019627, Part of Failsafe Collimator for VCT, ms13452 051015 This part is direct replacement for 5129998, and 5123185. ms13452