GE Healthcare · MRSigna
Head coil cables are damaged.
Symptom
MRIQ Grainy low SNR Quad Head Coil Images with Normal Prescan Values. Normal TG prescan values between 70 - 80. Extremity knee coil image are fine which uses the same xmt/rcv ports and indicates an issue isolated to the head coil.
Applies to
- Models
- Signa
- Product
- Signa LX Signa LX Product can be used with 8.x and 9.x systems.
- Software
- 9.1 For Signa LX product only
- Component
- 46-282441P2 Quad Head Coil cable white ends 46-282441P1 BLUE ENDS CABLE 46–282441P1 BLUE ENDS CABLE 46–282441P2 WHITE ENDS CABLE 46-282441P2 Quad Head Coil cable white ends
- Area
- MR
Full cause & step-by-step resolution