Communications to TCU, Z2 shim and coil ID failed due to defective SCSI card.
Symptom
TCU Communication Failures Coil ID Fails To Work And Z2 Shim Supply Not Responding. HFO system running HFO2 s/w with an Octane II computer. Unable to communicate with the TCU, also coil id is not functioning. When they plug the head coil into the MSHS it stays "RED", indicating coil ID failed. The thing to note here is that on the OpenSpeed, the serial line for the coil ID and the serial line for the TCU and the Z2 shim supply goes out through the SCSI Serial Expansion box. Did an scsistat and find the system does not identify the SCSI expansion box. HFO system running HFO2 s/w with an Octane II computer. Unable to communicate with the TCU, also coil id is not functioning. When they plug the head coil into the MSHS it stays "RED", indicating coil ID failed. The thing to note here is that on the OpenSpeed, the serial line for the coil ID and the serial line for the TCU and the Z2 shim supply goes out through the SCSI Serial Expansion box. Did an scsistat and find the system does not identify the SCSI expansion box.
Error message
2219981 Z2 supply is not up or not responding, Retrying Mon Dec 1 10:58:58 2003 2219981 1 hvmr port_mgr PMMain.cc 552 Z2 supply is not up or not responding, Retrying
Applies to
- Product
- HF OpenSpeed HFO1 HFO2 and HFO3 Only.
- Software
- HFO3
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Clarify Z2 Supply Is Not Up Or Not Responding. 3T Eclipse. 2219981 Z2 supply is not up or not responding, Retrying
- Need the Connections for the new Z2 Supply on the HFO
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- OrthoMRI Network cable is unplugged
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- CWA ID100060 Unable To Film.