GE Healthcare · CTLightSpeedVCT260134706430000999361
85381 LightSpeed 7.X
Symptom
Elevation Down will not work but elevation up does Elevation Down will not work but elevation up does. No table Touch Sensors Errors reported.
Error message
Error: 260134706 Elevation could not move with motion command. Check valve, inveter, or drive enable line from gantry. See Log Description for complete error Sat Dec 10 18:40:12 2005 Error: 260134706 Host: tgp Process: tgp_subsys File: gtcb_emc_movechk.cxx Line: 741 Sat Dec 10 18:40:12 2005 Host : Table Sub-System Ermes # : 260134706 Exception Class : Secondary Severity : Pri/Soft File : gtcb_emc_movechk.cxx Line# : 741 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 Elevation could not move with motion command. Check valve, inveter, or drive enable line from gantry. target position = 430000 um current position = 999361 um time from start = 1220 ms
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
- Component
- 5127285 SHUT OFF VALVE ASSY used on GT1700 and GT2000 table See Long Description 5127285 SHUT OFF VALVE ASSY used on GT1700 and GT2000 table Part #: 5127285 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: SHUT OFF VALVE ASSY Positoning GT FRU: Yes Status: Orderable Comments: Not Available
- Area
- CT
Full cause & step-by-step resolution