GE Healthcare · Nuclear MedicineMillennium212010832

During Full EDI upgrade, or any time computer is replaced the motion controller program is lost..

Symptom

During Full EDI upgrade, or any time computer is replaced the motion controller program is lost.. If gantry positon is in Cardiac mode, upon boot up, the gantry will not charactize the Radial axis During Full EDI upgrade, or any time computer is replaced the motion controller program is lost.. If gantry positon is in Cardiac mode, upon boot up, the gantry will not charactize the Radial axis. This requires the gantry to be manually cranked to Opposed mode. Contributor 212010832 Oheim September 21, 2007

Applies to

Models
Millennium
Product
MG/MC/Myosight Millennium MG/MC Millennium MyoSIGHT
Software
GenieAcq R4.5 V14 (M4) Genieacq 4.5 Build 14 (M4) 5220379 Genieacq SW R4.5 V014 M4 with accumulative patches GA450-05-1, GA450-6-1, GA450-07-0, GA450-08-1, and GA450-09-0 for Spare Part GENIE Acq R 4.5 V014 supports EDI and I/N board hardware configurations. The following computer platforms are supported by this version: SuperMicro P4SBR ( only supports the EDI detctor, reference SN43628 ) IBM x205 Dell 1400SC Dell GX200 Dell 1300. Dell 2200 and Intel AltServer are not supported. Reference CIS 5150522-100_r2
Component
*Gantry Gantry Gantry
Area
Nuclear Medicine

Full cause & step-by-step resolution