Monitor ordered for an Indigo system...no gamma sw received.
Symptom
NEC 1990Sxi Monitor received for an 8.3 Indigo system. Is it compatible? LX systems with Indigo Computer can only operate with 8.3 or 8.4 software. These systems originally shipped with CRT monitor. Ran out of stock of 2197655 FRU CRT monitor and FE Assist comments now state, as of 11/13/2006, to instead obtain 5117545 LCD monitor as it is a direct replacement. The problem with this is that, according to Software Engineering, all LCD monitors require gamma table software. No gamma table software supplied with this monitor. See long description for full details... Problem Description: LX systems with Indigo Computer can only operate with 8.3 or 8.4 software. These systems originally shipped with CRT monitor. Ran out of stock of 2197655 FRU CRT monitor and FE Assist comments now state, as of 11/13/2006, to instead obtain 5117545 LCD monitor as it is a direct replacement. The problem with this is that, according to Software Engineering, all LCD monitors require gamma table software. No gamma table software supplied with this monitor. Lack of gamma table software may result in perceived loss of IQ when viewing images from the LCD monitor. Need Engineering to determine if LCD monitor substitution poses any risks of perceived IQ degradation.
Applies to
- Models
- Signa
- Product
- Signa LX Signa LX Product can be used with 8.x and 9.x systems.
- Software
- 8.3 8.3
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.
- CWA ID100315. Auto and Manual Prescan Fails.
- What is the Acoustic Noise Level in the Ovation Magnet.