GE Healthcare · CTLightSpeed182904182924
LightSpeed 3.X (Ultra, Plus, QX/i) - Octane
Symptom
Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan
Error message
Error: 182904 A table side key is either stuck or pressed during initialization See long description for full error Tue Sep 11 09:36:32 2007 Error: 182904 Host: ETC Process: Table Controls File: tc_exec.c 1.27 1/15/96 Line: 475 Function: No System Function Reported Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Most Severe Time: 09:36:27:725 Log Series: 27 A table side key is either stuck or pressed during initialization or just after Emergency Off released. Port value 0x0100. Error: 182924 Up Footpedal (LOAD) is non-functional See long description for full error Tue Sep 11 09:36:32 2007 Error: 182924 Host: ETC Process: Table Controls File: tc_exec.c 1.27 1/15/96 Line: 512 Function: No System Function Reported Scan: 0/0/0 Type: None/Unknown Exception Level: Sec/Soft Time: 09:36:27:765 Log Series: 27 Up Footpedal (LOAD) is non-functional.
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 3.X (Ultra, Plus, QX/i) - Octane Always 4 or 8 slice MDAS 308i.2_H3.1M5 LightSpeed 3.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 308I.2_H3.1M5 305.4_H3.1M4 - obsolete 303.1_H3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2329241 FOOT PEDAL See long description for details Part #: 2329241 Part Type: GP CONSUMABLES Part Weight: 3.97 LBR Repairable: No Returnable: No Hazardous Material: N Description: FOOT PENAL FRU: Yes Status: Orderable Comments: FLS - FOOT PEDAL. ADDED PER GCN CT- TABLE-008 FOR THE "3 INJECTIONS PARTS" TRANSFER. THIS LCC PART REPLACES HCC PART NUMBER 46-296162P2. MS20544 ON 3/25/03.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.