LightSpeed 5.X (80/100kW)

Symptom

unable to scan after x-ray tube replacement with inverter max error .

Error message

Error: 260118159 (27-0907)Filament aging saturated. Z6,Z7= 0h,0h See LD for complete error Wed Mar 8 04:38:44 2006 Error: 260118159 Host: orp Process: orp_subsys File: gen_exec.c Line: 2150 Wed Mar 8 04:38:43 2006 Host : Jedi Ermes # : 260118159 Exception Class : Soft Severity : Pri/Soft File : gen_exec.c Line# : 2150 Function : No System Function Reported Scan Type : Axial Scan: 50262/5/48 (27-0907)Filament aging saturated. Z6,Z7= 0h,0h

Error: 260118211 ( 60-0313 )Inverter max. ILR current detected. This is a hardware detection of maximum current in serial resonant circuit. Z6,Z7= 0h,0h See long description for complete error Mon Feb 13 17:22:21 2006 Error: 260118211 Host: orp Process: orp_subsys File: gen_exec.c Line: 2150 Mon Feb 13 17:22:20 2006 Host : Jedi Ermes # : 260118211 Exception Class : Abort Severity : Primary File : gen_exec.c Line# : 2150 Function : No System Function Reported Scan Type : Static Scan: 50018/1/4 (60-0313)Inverter max. ILR current detected. This is a hardware detection of maximum current in serial resonant circuit.

Applies to

Models
LightSpeed, VCT, BrightSpeed
Product
LightSpeed 5.X (80/100kW) LightSpeed 5.X (80/100kW) ~ Use with both MDAS and GDAS ~ GRE Console HP_H_G16 HP_H_M16 HP_O_M16 LightSpeed 5.X (100kW) GDAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required