Pular para o conteúdo principal
GE Healthcare · MRSigna200002368200002352

LX System fails to boot applications Due to PCI Card Cage.

Symptom

Applications Software Will Not Boot.

Error message

200002368 A set of processes are deadlocked waiting for each other to start. Mon Apr 12 11:43:42 2004 Error: 200002368< Host: BaptistGEMR Process: hostMonitorbsp File: ssProcStart.c Line: 1099 A set of processes are deadlocked waiting for each other to start. svat, scn, rtcProxy, resetServer, lais, hgp, displayServer, autoview, app, WLServer, TIR, SITT, RtpHost, RTPSIServer, RIR, PPSproxy are the likely culprits. This deadlock has caused the system to shutdown!

200002352 A HOST_SHUTDOWN was initiated by the death of spectro. Mon Apr 12 11:43:32 2004 Error: 200002352 Host: BaptistGEMR Process: hostMoni tor File: ssProcStop.c Line: 810 A HOST_SHUTDOWN was initiated by the death of spectro.

9001 Unable to set connection. server NETserver host Mon Apr 12 11:43:35 2004 Error: 9001 Host: BaptistGEMR Process: protocol File: application.c Line: 274 Unable to set connection - server NETserver host BaptistGEMR.

Applies to

Models
Signa
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only
Component
2169940-53 PCI Card Cage For The Octane Computer
Area
MR

Full cause & step-by-step resolution