Scanner Hardware Reset Failed
Symptom
Scanner Hardware Reset Failed. See LD Scanner Hardware Reset Failed Hardware Failed to reset message appeared on monitor AFTER boot-up WITH Error:178568 Failed to establish socket connection to STC in 10 seconds and errors 244553 and 244554 and power up self test fails A, B, and 9 on STC CPU board The following errors were in the gesyslog Thu Jan 13 04:18:32 2005 Error: 178568 Suite: RP25 Host: SBC0 Process: fwmgr File: NBSSocket.m 1.9 Method: connectToTarget Line: 315 Failed to establish socket connection to STC in 10 seconds Thu Jan 13 04:18:32 2005 Error: 244553 Suite: RP25 Host: SBC0 Process: fwmgr File: HwSetUp.m 1.18 Method: hardReset Line: 429 The STC controller has failed to reset. Thu Jan 13 04:18:32 2005 Error: 244554 Suite: RP25 Host: SBC0 Process: fwmgr File: HwSetUp.m 1.18 Method: hardReset Line: 442 The reset failure of the OBC may be due to the reset failure of the STC. Also saw failures A, B, and 9 on the power-up test on the STC CPU .
Error message
Error : 230023002 Scanner Hardware is not ready or unavailable See LD for more info Fri Jun 10 08:22:33 2005 230023002 4 cr_ct_01 scanUtilsRx DetailedCalUtils.c 1160 FAILED to execute the Scan Request. Possible Reasons: Scanner Hardware is not ready or unavailable. Please verify and try again. *** Laser indicator flashing on the Gantry Display board – this indicates the system has lost communication with the ORP board. *** Additional error info: Error: 244640 HARD RESET through serial connection failed. some controllers(TGP, ORP) do not response to ping. Error: 244634 Immediate Request cannot be honored since the firmware is in the DOWN state Fri Jun 10 08:22:13 2005 230023002 host: cr_ct_01 process: scanUtilsRx DetailedCalUtils.c 1160 FAILED to execute the Scan Request. Possible Reasons: Scanner Hardware is not ready or unavailable. Please verify and try again. Fri Jun 10 08:22:33 2005 230023002 4 cr_ct_01 scanUtilsRx DetailedCalUtils.c 1160 FAILED to execute the Scan Request. Possible Reasons: Scanner Hardware is not ready or unavailable. Please verify and try again. Fri Jun 10 08:22:35 2005 244640 2 cr_ct_01 fwmgr fwmgr_fw_if.cxx 369 HARD RESET through serial connection failed.
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 5.X (60kW) LightSpeed 5.X (60kW) ~ Use with MDAS or GDAS HP_P_G4 HP_P_G8 HP_P_G16 HP_P_M16 LightSpeed 5.X 60KW G-DAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2218570-2 H-POWER FUSE KIT used on Lightspeed 5x systems. See Long Description for contents of fuse kit. Part #: 2218570-2 Part Type: GP CONSUMABLES Part Weight: 43.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: H-POWER FUSE KIT FRU: Yes Status: Orderable Comments: Not Available Fuse KIT contents Mark Number Name Qty Description 1 2351492 2.0 200A FUSE 2 2364059 2.0 GLASS FUSE 3 46-170021P50 2.0 "FUSE 12 AMPS 250 VOLTS BUSSMAN MDA12 DUAL ELEMENT." 4 46-170021P106 2.0 FUSE 8 AMPS 250 VOLTS BUSSMANN ONLY 5 2336517-2 2.0 FUSE 25 AMPS 700 VOLTS 2.5M SECONDS 6 46-170021P52 3.0 "FUSE 3 AMPS 250 VOLTS(REVIEWED TK, 7/91)." 7 46-170021P10 2.0 "006.000A 0250V 3AG FAST UL/C 8 46-170021P101 1.0 "FUSE 20 AMPS 700 VOLTS 2~ LONG X 9/16~ DIA FAST BLOW" 9 46-170021P11 2.0 "7.0A, 250V SLO-BLO FUSE. *TYPE 3AG, 1.25~ X 0.25~ GLASS BODY. 10 46-170021P31 2.0 "1/2A, 250V SLO-BLO FUSE. TYPE 3AG, 1.25~ X 0.25~ GLASS BODY. 11 2106993-5 2.0 FUSE 20 AMPS 500 VOLTS .3 SECONDS 12 46-327160P1 2.0 "12.0A, 125VAC, DUAL ELEMENT TIME DLY FNM-12, 0.406 X 1.5 LG, FIBER BODY" 13 2379651 1.0 FUSE - 700 V, 200 A, FAST SEMI
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.