GE Healthcare · CTLightSpeed26000793750444
Fastcal fails alpha calibration
Symptom
Scan abort during alpha calibration. Scan abort during alpha calibration
Error message
Error: 260007937 DAS Converter Board(s) encountered a stage overflow during alpha calibration. See Long Description for full error. Tue Dec 28 09:41:32 2004 Error: 260007937 Host: orp Process: orp_subsys File: brdflt.c Line: 82 Tue Dec 28 09:41:31 2004 Host : Das Ermes # : 260007937 Exception Class : Soft Severity : Sec/Soft File : brdflt.c Line# : 82 Function : No System Function Reported Scan Type : None/Unknown Scan: 50444/1/0 DAS Converter Board(s) encountered a stage overflow during alpha calibration. See Following Error Message for Converter Card List Board : 87 ASIC/Channel : 0x012c Stage : 1 Board Channel : 41
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 5.X (60kW) LightSpeed 5.X (60kW) ~ Use with MDAS or GDAS HP_P_G4 HP_P_G8 HP_P_G16 HP_P_M16 LightSpeed 5.X 60KW G-DAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2371627-4 replaces 2371627-3 GDAS / MDAS 16 Slice RIGHT Backplane Chassis Asm EDITION 1-2 See LD for FEMC comments 2371627-4 replaces 2371627-3 GDAS / MDAS 16 Slice RIGHT Backplane Chassis Asm EDITION 1-2 Part #: 2371627-4 Part Type: GP CONSUMABLES Part Weight: 2.18 LBR Repairable: No Returnable: No Hazardous Material: N Description: GDAS-MDAS 16 Slice RIGHT BP-Chassis Asm EDITION 1-2 FRU: Yes Status: Orderable Comments: Added as replacement for 2371627, -2, -3 & 2332630 per ecr 2012678, ms13452 050518 Part #: 2371627-3 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: BP-Chassis Right EMC Asm - GDAS FRU: Yes Status: Orderable Comments: 2371627-3 is backward compatible replacement for 2371627-2. DO NOT USE 2371627-2 to replace 2371627-3. 2371627-3 is EMC2 Compliant. ms13452 PART ADDED PER VCT RSPL AND IS EMC2 COMPLETE MS8336 12/08/2004
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.