5X system crashes upon applications boot attempt due to failure of the "display" process.
Symptom
Genesis System Crashes Upon Applications Boot Attempt With Failure of the "Display" Process See long description The system is a 5X running 5.8 M4 software. The system fails to boot up applications, it will make it up to the point of “you have 10 seconds to stop autostart” which means Unix OS has come up. Before the software crashes the following error is displayed: “Tempmon.dat read error, continuing without check for abnormal shutdown. Taking value for irated 0.000000. Taking value for time constant 0.000000” If we stop autoboot at unix level, we can take a look at the message log. There are a number of errors that are related to the display process. Errors in the log state that the display process was what caused the host to shutdown. Additionally there is a core.display created on every bootup attempt. The FE has tried to replace many parts including: The Stealth CPU, GIP, GAI, GFB, and the GCP board. Also reloaded software with no difference in the issue.
Error message
2218632 Temp Monitor, Library Call Failure, .. Read/Write Error Shutting down local host | due to startup failure: display Time expired for startup: display in 300 seconds emergency shutdown: display
Applies to
- Models
- Signa
- Product
- Signa Advantage Includes 4.x and 5.x systems. The software that this solution is tied to is indicated by the Software concept below.
- Software
- 5.X
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.