Prestige - Error M3410, DC Bus Out of Range
Symptom
Prestige with GPX console gives error M3410, Bux out of range. Check fuses F1, F2, F3, F4 & F5 on DC Filter board MPH A3 A1 (p/n 2103601). At least one of them is open. Prestige with GPX console gives error M3410. Check fuses F1, F2, F3, F4 & F5 on DC Filter board MPH A3 A1 (p/n 2103601). At least one of them is open.
Error message
Error M3410, DC Bus Out of Range
Applies to
- Product
- Prestige VI Prestige VH MPH Prestige II Prestige SI
- Component
- GPX
- Area
- X-ray R&F
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare X-ray R&F issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required