Ventri system, no counts or image on detector when setup for acquisition.

Symptom

Ventri system, no counts or image on detector when setup for acquisition.

Error message

Adam Test failure :Analog Window calculated energy is out of nominal bounds : Could not create analog energy window Log from Adam test diaganostic with errors- Adam Test failure- 09:07:21 Testing detector # 2 09:07:27 Calculated start DC slide for Analog Window: 0x750 09:07:27 Calculated end DC slide for Analog Window: 0x7f0 09:07:27 Calculated G0 for analog channel: 0x712 09:07:27 Calculated Energy for Analog Window: 0x7b2 27/07/07 09:07:27 Analog Window calculated energy is out of nominal bounds - Min: 0x300, Max: 0x500 09:07:27 Forcing Enr value to 0x400 09:07:27 Could not create analog energy window

Applies to

Product
Ventri Ventri Ventri
Component
UGP500114 (5272330) ADAM Board (Ventri, Infinia, VG) FE Assist: If PCA000464 or UGP500114 are not available order # 5272330 UGP500114 (5272330) ADAM Board (Ventri, Infinia, VG) FE Assist: If PCA000464 or UGP500114 are not available order # 5272330 *Detector (Imaging Detector) *Detector (Imaging Detector) *Detector (Imaging Detector)