Pular para o conteúdo principal
GE Healthcare · CTLightSpeedVCT245207200107002200107023

LightSpeed 4.X - GRE

Symptom

Invalid or mistached DAS type messages appear when tube warm up is attempted. Invalid or mistached DAS type messages appear when tube warm up is attempted

Error message

Error: 245207 Software has detected an invalid or mistached DAS type. This will prevent the system from scanning. see long description Mon Apr 3 08:29:12 2006 Error: 245207 ls16 scanmgr validate request.cxx 378 Software has detected an invalid or mistached DAS type. This will prevent the system from scanning.

Error: 200107002 Invalid event 1014 (SCAN DONE event) for current state 2000 (Idle). See Long Description for complete error. Wed Nov 22 08:55:38 2006 Error: 200107002 Host: ct01 Process: scanRx File: srx_sm_event_notify.c Line: 703 Function: Data Acquisition : OC Processing Invalid event 1014 (SCAN DONE event) for current state 2000 (Idle).

Error: 200107023 A Failure had been detected while calling UIR x_ToScanRequest function or srx_sm_generate_action function See Long Description for error variations Tue Oct 19 09:22:20 2004 Error: 200107023 Host: dgct Process: scanRx File: srx_sma_manage_scan.c Line: 688 Function: Data Acquisition : OC Processing srx_sma_prepare_send_request: A Failure had been detected while calling UIRx_ToScanRequest function. Nov 22 08:55:38 2006 Error: 200107023 Host: ct01 Process: scanRx File: srx_sm_event_notify.c Line: 148 Function: Data Acquisiti

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 4.X - GRE See Long Description for detailed information - H2_P_M16 Product Name: LightSpeed 4.X with GRE Product Photographs GE HEALTHCARE Method of remote or software identification - See attachment below for a complete list of software commands and typical results 1. check_config 2 . swhwinfo “swhwinfo” will have the following switches: · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console types: GOC3 (GRE) GOC4 (GRE) List of software the product can run: 07MW11.10 SP2.0, SP3.0, SP3.1 FAQ's Table weight limit - 180 kg (400 lbs) normal operation - 205 kg (450 lbs) maximum allowed with decreased accuracy Tube type - D3186T D3186T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.