Symptom

Tube identification hardware not detected reported on boot up

Error message

Error: 260118314 Could not verify tube ID against target configuration Status code: -3 When the system can not identify the tube the system will post a Pop-up window on every boot up, but will not stop the customer from scanning. Error 260118314 5 orp orp_subsys gen_exec.c 3695 No System Function Reported Soft Severity : Pri/Soft None/Unknown Scan: 0/0/0 Could not verify tube ID against target configuration Status code: -3

Error: 244637 System could not find the tube type auto-detect device. Scanning is not disabled. However, please verify the tube type. See LD for complete error Mon Nov 15 13:41:38 2004 Error: 244637 Host: ct01 Process: fwmgr File: fwmgr_fw_if.cxx Line: 441 Function: No system function reported Scan Type: Not scanning Scan: 0/0/0 Scan Seq Id: 0 System could not find the tube type auto-detect device. Scanning is not disabled. However, please verify the tube type.

Error: 244638 Scanning with the improper tube configuration can cause damage to the tube. Continue at your own risk Status from the generator: -4 See LD for complete error Mon Nov 15 13:41:38 2004 Error: 244638 Host: ct01 Process: fwmgr File: fwmgr_fw_if.cxx Line: 443 Function: No system function reporte

Applies to

Models
LightSpeed
Product
LightSpeed 5.X (80/100kW) LightSpeed 5.X (80/100kW) ~ Use with both MDAS and GDAS ~ GRE Console HP_H_G16 HP_H_M16 HP_O_M16 LightSpeed 5.X (100kW) GDAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.