Pular para o conteúdo principal
GE Healthcare · MRSigna11002110331102411029

Clarify Unknown Modality Type Errors Transferring Images From Workstation

Symptom

Unable To Transfer (Pull) Images To Scanner Customer can query an ALL image store system from the LX console, but transfers (get) fail.

Error message

11002 DCM low level error. Network error (508). Network error (114) Thu Sep 30 07:56:30 2004 Error: 11002 Host: chmr Process: dcserver File: dcscommon.c Line: 228 DCM low level error: Network error (508): Network error (114)

11033 Unknown modality type US Mon Nov 5 11:38:32 2001 Error: 11033 Host: MRS1 Process: dcserver File: dcsformat.c Line: 1671 Unknown modality type US

11024 Failure to parse FIND response: Continuing! Mon Nov 5 11:38:32 2001 Error: 11024 Host: MRS1 Process: dcserver File: dcslist.c Line: 511 Failure to parse FIND response: Continuing!

11029 MFM Warning: Error:element is not UI UID string Mon Nov 5 11:38:58 2001 Error: 11029 Host: MRS1 Process: dcserver File: dcsformat.c Line: 1715 MFM Warning: Error:element is not UI (UID string) (-102) Group:0x8 Element:0x18

Applies to

Models
Signa
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required