Prestige VH - JACK 1 Plastic spacers are grinding away.
Symptom
Prestige VH - JACK 1 Plastic spacers are grinding away. Suspect wrong thickness of spacer used. Jack 1, where the jack connects to the table, has plastic washers. These washers or spacers are grinding away. Suspect the wrong thickness of spacers were used in this system. If the wrong size spacer is used it can cause the jack not to be perpendicular to the table. This can cause friction on the spacers resulting in premature wear.
Applies to
- Product
- Prestige VH
- Area
- X-ray R&F
Full cause & step-by-step resolution