Prestige SI GPX - Rapid sequence 3rd format position dark exposure.
Symptom
Prestige SI GPX - Rapid sequence 3rd format position dark exposure. Prestige SI GPX System: Rapid sequence selected, 3rd exposure is dark. Only happens when rapid sequence is selected. Trouble shooting led to monitoring the AEC ramp signal. This showed that the 3rd exposure is longer then exposures of the other positions.