Prestige VH - No medium power servo movement.

Symptom

Prestige VH - No medium power servo movement. Prestige VH no SID, table top, and palpator movement. High and low servo motion is allowed. Checked fuses 3F3 and 3F1, both are good. Next checked fuses 2F50 and 2F49. 2F49 was found to be bad. All these fuses supply 32 volts to medium servo drive boards.

Applies to

Product
Prestige VH
Area
X-ray R&F