Pular para o conteúdo principal
GE Healthcare · MRSigna2245096

SGD Gradient Driver Static Test Fails due to faulty PPBM.

Symptom

ACGD Gradient Driver Static Test Fails ACGD Gradient Driver Test Fails

Error message

2245096 Gradient driver test - Statc test failed on X axis, VRAIL had 3 failures out of 3. Possible causes on this axis are GPM, GPM PS or GIP. Static Standby mode 2245096 Gradient driver test - Statc test failed on X axis, VRAIL had 3 failures out of 3. Possible causes on this axis are GPM, GPM PS or GIP. Static Standby mode

Applies to

Models
Signa
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only
Component
2175599-2 SGD Hi Slew GPM PPBM module
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.