Compax40E - Service Note 4106A Fragile Bucky Handle Mounting
Symptom
Compax40E - Service Note 4106A Bucky Handle Mounting is to Fragile. 1. For all Compax tables, the Bucky Handle is too fragile and is easily broken. 2. The former Bucky Handle used for all Compax 40E is not compatible with the wiring for the Compax 40E (no CE). For the Bucky Handle of the Compax 40E (no CE) the enable push-button switch is in an open configuration different from the close configuration of the Compax 40E (CE). 3. Some Bucky Handle are not delivered with the minimum screw length. This causes the Handle to break easily.
Applies to
- Component
- Compax 400T
- Area
- X-ray R&F