Symptom

The system would boot to applications successfully but failed to perform a scan. Error message received when the system set up to perform a scan. The system would boot to applications successfully but failed to perform a scan. Open LD for error log string The system would boot to applications successfully but failed to perform a scan. Error message received when the system set up to perform a scan. Errors 200001830 and 245181 Start: Prospective Exam: 483 : Protocol: 10.1 1120236674 0 1 Fri Jul 1 16:51:14 2005 200001830 4 ct02 scanmgr ecommL_func.c 106 IPC failure caused by cupIpcConnect from scan_mgr to dataacq. 1120236674 1 1 Fri Jul 1 16:51:14 2005 245181 4 ct02 scanmgr dataacq_if.cxx 599 Function: System Control : Scan Control Scan Type: Not scanning Scan: 50142/1/0 Scan Seq Id: 2 Unable to send ecomm event. Event: EV_DATAACQ_CANCEL_ALL_ACQS (3420) Destination: dataacq IPC failure.

Error message

Error: 245048 The dataacq process appears to be down - error returned is -1 See long description for full error Fri Dec 3 14:55:36 2004 Error: 245048 Host: ct99 Process: scanmgr File: start_finish_request.cxx Line: 282 Function: System Control : Scan Control Scan Type: Not scanning Scan: 50000/1/0 Scan Seq Id: 3 The dataacq process appears to be down - error returned is -1

Error: 245181 IPC failure Unable to send ecomm event See long description for full error Fri Dec 3 14:55:36 2004 Error: 245181 Host: ct99 Process: scanmgr File: dataacq_if.cxx Line: 505 Function: System Control : Scan Control Scan Type: Not scanning Scan: 50000/1/0 Scan Seq Id: 3 Unable to send ecomm event. Event: EV_DATAACQ_CANCEL_ALL_ACQS (3420) Destination: dataacq IPC failure.

Error: 245050 Error verifying that Data Acquisition is ready for scanning Open Long Description to see error examples Error: 245050 Error verifying the Data Acquisition is ready for scanning. More than one error text exists for this error below: Fri Jul 1 16:51:14 2005 Error: 245050 Host: ct02 Process: scanmgr File: start_scanning.cxx Line: 231 Function: System Control : Scan Control Scan Type: Not scanning Scan: 483/1/0 Scan Seq Id: 

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
Component
5114523-2 Westville VDARC - Computer, Data Acquisition Recon Control (equivalent to 5114523) See long description for details Part #: 5114523-2 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 2.43 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: Computer, Data Acquisition Recon Control - Ed 2 FRU: Yes Status: Orderable Comments: **FE Assist** applies to CT Not Pet systems. If no stock is available order a Jarrell Darc 5159833 to replace a Westville, you MUST also order p/n 5193989 software release 06MW29.7 or LATER , and perform a LFC. See Service note SN06101010A.(2/12/07 212001261) ** Second Opinion ** Reload darc OS & APPS. For reasons unknown darc SW gets corrupt. DARC failures in system that are caused by noisy p/s fans should have a PQR written to part. IP network address should not be changed unless hospital network is on the same IP subnet. Make sure the proper DARC is ordered to match DAS/Slipring. Possible reload of DARC may fix issues. Added per ECR 2016469, this part is forward backward compatible replacement for 5114523-2. Do not use 5114523 to replace 5114523-2. ms13452 050921 Part #: 5114523 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: Computer, Data Acquisition Recon Control - VDARC FRU: Yes Status: Orderable
Area
CT

Full cause & step-by-step resolution