OrthoMRI Foot/Ankle cable coil is broken.
Symptom
OrthoMRI 18008 w: frequency calibration during scout, there is no signal message.
Error message
OrthoMRI 18008 w: frequency calibration during scout, There is no signal OrthoMRI 18008 w: frequency calibration during scout. There is no signal.
Applies to
- Product
- OrthoMRI Escan XQ EscanXQ and EscanNE
- Software
- OrthoMRI S/W Release 9.1B. OrthoMRI S/W Release 9.1B.
- Component
- OrthoMRI P/N LNR8800031002 CABLE FOR FOOT/ANKLE COIL. OrthoMRI P/N LNR8800031002 CABLE FOR FOOT/ANKLE COIL. There is a PSDB solution available for deselecting the automatic coil recognition for ALL OrthoMRI systems. If need be, follow that resolution. Then the system can be used as normal once the operator manually selects which coil is being used while on the patient set up page. The most common failure is the particular coil's cable. If multiple coils have this issue the fix could be the mechanical swing arm.
- Area
- MR