Fuji CRAC3 CRAC3CS CRAC3CSID CR9000 CR9501

Symptom

Fuji CRAC3 CRAC3CS CRAC3CSID CR9000 CR9501: Power supply AC supply input voltage config. On Fuji units the power supply can be configured for 100, 116, 230 or 240 AC input voltage depending on the site. The power supply is configured by a jumper plug that plugs into CN4 of the power supply. The instructions or schematic for each configuration of the plug are not very easy to find in the Fuji service manual.

Applies to

Component
FUJI
Area
X-ray R&F

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare X-ray R&F issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required