Tilt C - Error 4AF1, 4A93, 4A8E and 4A51

Symptom

Tilt C - Errors: 4AF1, 4A93, 4A8E and 4A51 Tilt-C 4AF1 Error. FE found that at test point 19 the 5.1 volts off of CR31 was not 5.1v it was 4.7v. If you look on the VIS1 board (15v supply, CP1A11A7(E1H1))and check TP19, which should be +5v. shown on page 7-95 of direction 36003623.

Error message

Tilt C Error 4A51

Tilt C Error 4AF1

Tilt C Error 4A83

Tilt C Error 4A93