Tilt C - Error 4AF1, 4A93, 4A8E and 4A51
Symptom
Tilt C - Errors: 4AF1, 4A93, 4A8E and 4A51 Tilt-C 4AF1 Error. FE found that at test point 19 the 5.1 volts off of CR31 was not 5.1v it was 4.7v. If you look on the VIS1 board (15v supply, CP1A11A7(E1H1))and check TP19, which should be +5v. shown on page 7-95 of direction 36003623.
Error message
Tilt C Error 4A51 Tilt C Error 4AF1 Tilt C Error 4A83 Tilt C Error 4A93