GE Healthcare · MRSigna123456
The Performed Procedure Step option was installed but not configured at the HIS Broker which caused this issue
Symptom
When Exams are networked to the PACS, each Series shows as a separate Exam For example, Exam 123456 has 7 series in it. When networked to the PACS, instead of having 7 series in the one exam, shows as 7 separate exams, all with the same exam number.
Applies to
- Models
- Signa
- Product
- Signa LX Signa LX Product can be used with 8.x and 9.x systems.
- Software
- 9.1 For Signa LX product only
- Area
- MR
Full cause & step-by-step resolution