PROTEUS Wallstand cable wiring error, no exposure.

Symptom

PROTEUS Manual Collimator to table worked fine. Shooting to the SG60 W/S, the UIF was blank when pointed to the W/S. The W/S Config was not to the correct end of the room, it was installed at the foot but configured for the head, thus the blank display. Once the W/S was correctly configured, still could not make exposures.

Applies to

Product
PROTEUS XRa
Area
X-ray R&F