GE Healthcare · CTHiSpeedLightSpeedBrightSpeed185122182902
Table would not move in or out
Symptom
Table would not move in or out. Table would not move in or out
Error message
Error: 185122 24V Power Supply Out Of Tolerance Sun Sep 5 16:04:34 2004 Error: 185122 Suite: RP21 Host: ETC Process: Control File: etcPowSupTest.c 1.4 2/16/99 Method: No Method Line:117 Function: No System Function Reported Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Soft Time: 3588 Log Series: 2 HW error: 24V Power Supply Out Of Tolerance: A/D reading = 0x4 Calc Voltage = 0.03 (Volts) Tolerance = 3 (Percentage) INFO: If the supply is low the problem may be that the Xray Drive Power is OFF Error: 182902 Emergency Off button has been pressed Sun Sep 5 16:04:39 2004 Error: 182902 Suite: RP21 Host: ETC Process: Table Controls File: tc emergency.c 1.6 2/16/99 Method: No Method line: 89 Function: No System Function Reported Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Soft Time: 4016 Log Series: 22 Emergency Off button has been pressed
Applies to
- Models
- HiSpeed, LightSpeed, BrightSpeed
- Product
- HiSpeed QX/i - GOC2 103L.5_HSQXIM4 103L.4_HSQXIM4 102L.1_HSQXIM3 HiSpeed QX/i - GOC2 Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types GOC2 List of softwares the product can run 103L.5_HSQXIM4 103L.4_HSQXIM4 - obsolete 102L.1_HSQXIM3 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required