GE Healthcare · MROvationSigna2225554
Multicoil T/R Driver faults (always on port 4) only with the USAI Large Knee coil.
Symptom
Multicoil T/R Driver faults (always on port 4) only with the USAI Large Knee coil. Signa Ovation could not scan with the USAI Large Knee coil. All other coils worked fine. The coil was replaced but still the failure occurs.
Error message
2225554 Multi-Coil T/R Driver Board Fault. An Open Circuit was Detected on Port 4. Thu Aug 21 12:06:10 1997 Suite: MRS1 Host: OC0 Proc: HPC_PORT_ID Error: 2225554 File: PMH.m 1.29 Method: spi_tr_mon Line: 6260 Multi-Coil T/R Driver Board Fault. An open circuit was detected on port 4. Check cable to multi-coil board or check for an open pin diode.
Applies to
- Models
- Ovation, Signa
- Product
- Ovation/MFO
- Software
- MFO3 Applies to Ovation3 with HP4000 computer running Linux OS.
- Area
- MR
Full cause & step-by-step resolution