Defective Analogic RF amplifier.
Symptom
Unable to Scan with Analogic RF Amplifier Power Supply Faults 2238875 & 2238877 Unable to scan due to Analogic RF Amplifier Power Supply Faults as shown in Long Description Mon Jun 11 09:46:41 2007 Error: 2238875 Host: MYMR3 Process: NSP File: HPC_PORT_ID:spi_rf_mon Line: 3068 Analogic RF Amplifier Power Supply Fault. Analogic RF Amplifier indicates the main power supply is out of specification. Possible causes: Supply feedback loop exceeds specification, PS heat sink is overheated, Power supply FETS are shorted. Mon Jun 11 09:42:20 2007 Error: 2238877 Host: MYMR3 Process: NSP File: HPC_PORT_ID:spi_rf_mon Line: 3068 Analogic RF Amplifier I Peak Overloaded Fault. Line: 1621 Instantaneous peak drain current of RF devices exceeds specification.
Error message
2238877 Instantaneous peak drain current of RF devices exceeds specification. 2238877 Host: MYMR3 Process: NSP File: HPC_PORT_ID:spi_rf_mon Line: 3068 Analogic RF Amplifier I Peak Overloaded Fault. Line: 1621 Instantaneous peak drain current of RF devices exceeds specification. Error: 2238877 Host: MYMR3 Process: NSP File: HPC_PORT_ID:spi_rf_mon Line: 3068 Analogic RF Amplifier I Peak Overloaded Fault. Line: 1621 Instantaneous peak drain current of RF devices exceeds specification. 2238875 Analogic RF Amplifier Power Supply Fault. Analogic RF Amplifier indicates the main power supply is out of specification. Possible causes: Supply feedback loop exceeds specification, PS heat sink is overheated, Power supply FETS are shorted. 2238875 Host: MYMR3 Process: NSP File: HPC_PORT_ID:spi_rf_mon Line: 3068 Analogic RF Amplifier Power Supply Fault. Analogic RF Amplifier indicates the main power supply is out of specification. Possible causes: Supply feedback loop exceeds specification, PS heat sink is overheated, Power supply FETS are shorted. Error: 2238875 Host: MYMR3 Process: NSP File: HPC_PORT_ID:spi_rf_mon Line: 3068 Analogic RF Amplifier Power Supply Fault. Analogic RF Amplifier indic
Applies to
- Models
- Signa
- Product
- HF OpenSpeed HFO1 HFO2 and HFO3 Only.
- Software
- HFO3
- Component
- 2243079-2 0.7T RF POWER MODULE W/FRU CASE 0.7T RF POWER MODULE W/FRU CASE 2243079-2 0.7T RF POWER MODULE W/FRU CASE 2243079-2
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.