GE Healthcare · X-ray R&F
Bennett Silhouette SE - Errors
Symptom
Bennett Silhouette SE - Errors: SES 410 and KVP - C ERROR "SES 410" still happens after the replacement of new Source Board (P/N: 2132436). Found 1. Fuse F1 (1.5 A) blown. 2. R32 (2 Ohm-5 Watt) open circuit. 3. R3 (200 ohm-25 Watt) open circuit. 4. Diode D10 (1N4005) short circuit.
Error message
KVP-C Error KVP-C Error SES 410 Error
Applies to
- Product
- Silhouette HF
- Area
- X-ray R&F
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare X-ray R&F issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required