Bennett Silhouette SES 410 Error at Boot Up
Symptom
Silhouette SES 410 Error at Boot-Up SES error indicates "Stored Energy System" (battery voltage is out of spec) Can be caused by; Battery voltage too low or too high. (410 VDC +- 10%) The two banks of batteries associated with one source charger board are out of balance. (couldn't find spec) Or one bank out of spec. (nominally 204 to 206 VDC) Sense circuit adjust is out of cal.
Error message
SES 410 Error
Applies to
- Product
- Silhouette HF
- Area
- X-ray R&F
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare X-ray R&F issue are available to FSELIB subscribers — re-authored, web-first, searchable.