GE Healthcare · MRSigna2225703
Unable to Scan with MDS Link Error Code 2225703 Due to SIGNAL COLLECTING BOARD SCB.
Symptom
Unable to Scan with MDS Link Error Code 2225703 After Replacing the SCB Board The FE at this site replaced the SCB board (Signal Collecting Board) in the System cabinet for an unrelated reason. But after installing the replacement board he was unable to scan due to error code 2225703 "The MDS fiber optic link appears to be broken between the Gradient and the RF cabinets". TPS Reset does not help. PQR 13114612.
Error message
2225703 MDS link failure. The MDS fiber optic link appears to be broken between the Gradient and the RF cabinets. Check gradient cabinet power, fiber optic connections and signal level of fiber optic link. Run MDS fiber optic link diagnostics 2225703 MDS link failure. The MDS fiber optic link appears to be broken between the Gradient and the RF cabinets. Check gradient cabinet power, fiber optic connections and signal level of fiber optic link. Run MDS fiber optic link diagnostics
Applies to
- Models
- Signa
- Product
- Signa Horizon (5X Horizon) 5X Horizon systems
- Software
- 5.X
- Component
- 46-288166G1 SIGNAL COLLECTING BOARD SCB MR2 A23 A2
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.
- CWA ID100315. Auto and Manual Prescan Fails.
- What is the Acoustic Noise Level in the Ovation Magnet.