GE Healthcare · MRSigna2225703
Unable to Scan with MDS Link Error Code 2225703 Due to SIGNAL COLLECTING BOARD SCB.
Symptom
Unable to Scan with MDS Link Error Code 2225703 After Replacing the SCB Board The FE at this site replaced the SCB board (Signal Collecting Board) in the System cabinet for an unrelated reason. But after installing the replacement board he was unable to scan due to error code 2225703 "The MDS fiber optic link appears to be broken between the Gradient and the RF cabinets". TPS Reset does not help. PQR 13114612.
Error message
2225703 MDS link failure. The MDS fiber optic link appears to be broken between the Gradient and the RF cabinets. Check gradient cabinet power, fiber optic connections and signal level of fiber optic link. Run MDS fiber optic link diagnostics 2225703 MDS link failure. The MDS fiber optic link appears to be broken between the Gradient and the RF cabinets. Check gradient cabinet power, fiber optic connections and signal level of fiber optic link. Run MDS fiber optic link diagnostics
Applies to
- Models
- Signa
- Product
- Signa Horizon (5X Horizon) 5X Horizon systems
- Software
- 5.X
- Component
- 46-288166G1 SIGNAL COLLECTING BOARD SCB MR2 A23 A2
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.