Pular para o conteúdo principal
GE Healthcare · MRSigna2225703

Unable to Scan with MDS Link Error Code 2225703 Due to SIGNAL COLLECTING BOARD SCB.

Symptom

Unable to Scan with MDS Link Error Code 2225703 After Replacing the SCB Board The FE at this site replaced the SCB board (Signal Collecting Board) in the System cabinet for an unrelated reason. But after installing the replacement board he was unable to scan due to error code 2225703 "The MDS fiber optic link appears to be broken between the Gradient and the RF cabinets". TPS Reset does not help. PQR 13114612.

Error message

2225703 MDS link failure. The MDS fiber optic link appears to be broken between the Gradient and the RF cabinets. Check gradient cabinet power, fiber optic connections and signal level of fiber optic link. Run MDS fiber optic link diagnostics 2225703 MDS link failure. The MDS fiber optic link appears to be broken between the Gradient and the RF cabinets. Check gradient cabinet power, fiber optic connections and signal level of fiber optic link. Run MDS fiber optic link diagnostics

Applies to

Models
Signa
Product
Signa Horizon (5X Horizon) 5X Horizon systems
Software
5.X
Component
46-288166G1 SIGNAL COLLECTING BOARD SCB MR2 A23 A2
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare MR issues