Pular para o conteúdo principal
GE Healthcare · MRSigna2254052

Wrong Patient Weight Entered for Phantom

Symptom

Unable to Scan a Phantom With RF Errors

Error message

2254052 Power Monitor 2 Fault. RF Transmit power has reached safety limit allowed for scan parameters and Patient Weight Fri Aug 20 08:54:29 2004 Error: 2254052 Host: lx-mr Process: NSP File: SCP:PmMdsHWO Power Monitor 2 Fault. RF Transmit power has reached safety limit allowed for scan parameters and patient weight. Check patient weight entered. amplitude: 35 pulse width: 33 duty cycle: 09 Fault in head amplitude. If you need to confirm the source of the fault, reset the power monitor and repeat the scan

Applies to

Models
Signa
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare MR issues