Pular para o conteúdo principal
GE Healthcare · CTLightSpeedVCT

LightSpeed 3.X (Ultra, Plus, QXi) - Octane

Symptom

Root Partition is almost full ( 98%) - customer is unaware of this condition and store log does not help. See Long Description for more info VCT Root Partition is almost full ( 98%) -Customer is unaware of this condition and store log does not help. Once the Root partion is full the System will reports file system is full to the customer. [root@qehct1]# df -k Filesystem 1K-blocks Used Available Use% Mounted on /dev/sdb2 4182720 3982208 200512 96% / /dev/sdb1 101086 4798 91069 6% /boot none 1944112 0 1944112 0% /dev/shm /dev/sdb9 9765248 272 9764976 1% /usr/interchange /dev/sdb10 11717120 792 11716328 1% /data /dev/sdb11 37670048 9633924 28036124 26% /usr/g /dev/md0 143363584 100004448 43359136 70% /usr/g/sdc_image_pool

Error message

Disk Space is Low message appears when rebooting saying , Call GE Service. See Long Description for details While scanning or booting up to applications, a Pop Up GUI appears saying: 1. Software Problem Detected on System. 2. Disk Space is Low. 3. See Error Log for which partition is causing the problem. 4. Complete current processes and shutdown system. 5. Call GE Service.

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 3.X (Ultra, Plus, QX/i) - Octane Always 4 or 8 slice MDAS 308i.2_H3.1M5 LightSpeed 3.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 308I.2_H3.1M5 305.4_H3.1M4 - obsolete 303.1_H3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Software
308I.2_H3.1M5 for LightSpeed 3.X (Ultra, Plus, QX/I with H3 Console Assembly (2266832-3, 2304732-3) or GOC1-Octane2 Console (2341104, -2) - showprods returns LightSpeed Apps ver 308I.2_H3.1M5 308I.2_H3.1M5 for LightSpeed 3.X (Ultra, Plus, QX/I with H3 Console Assembly (2266832-3, 2304732-3) or GOC1-Octane2 Console (2341104, -2) - showprods returns LightSpeed Apps ver 308I.2_H3.1M5 308I.2_H3.1M5 for LightSpeed 3.X (Ultra, Plus, QX/I with H3 Console Assembly (2266832-3, 2304732-3) or GOC1-Octane2 Console (2341104, -2) - showprods returns LightSpeed Apps ver 308I.2_H3.1M5 Class C Application Software Set CD Part No. Ver. CD Label CD Part No. Ver. CD Label Set Part No. Apps. CD Part No Apps. CD Label showprods/ Service Desktop Intro. Date Comment 2395172 1104I.2_ H4.0M5 CLASSC LightSpeed MDAS 4/8/16/ HiSpeed QXi Class C Software 2395177 2395162, 2395163, 2399867 LightSpeed Plus/ Ultra MDAS Apps Software (Disk 1, Disk 2 and Patch A) 308I.2_H3.1M5 FW47-03 LS2002M5. IRIX Only. Class C/M Common w/ LS 2X, 3X, & 4X.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues