GE Healthcare · CTHiSpeedVCT2300150241127205412245189524000230015035512005
Unable to perform scan at certain KV
Symptom
Unable to perform scan at certain KV stations
Error message
Error: 230015024 A corrupt file has been found in the cal database See LD for complete error Tue Sep 20 08:36:51 2005 Error: 230015024 Host: ls16 Process: scanmgr File: cal_db_file_io.c Line: 1983 Function: CAL Processing : SBC Processing A corrupt file has been found in the cal database. The corrupt file has been removed from the cal database and saved for later analysis. Corrupt File: /usr/g/cal_db/kv120/large_sfov/16x0625/large_spot/dasgain5/cal.1.acal Saved File: /usr/g/cal_db/recovery/kv120.large_sfov.16x0625.large_spot.das.cal.1.acal.1127205412 Error: 245189 A valid calibration module could not be derived from the calibration database. Invalid Calibration Vector: acal, Row: 0. See LD for complete error Tue Oct 11 10:45:52 2005 Error: 245189 Host: er_ct5_vct Process: scanmgr File: scanmgr_lib.cxx Line: 229 Function: System Control : Scan Control Scan Type: Axial Scan: 0/0/0 Scan Seq Id: 524000 A valid calibration module could not be derived from the calibration database. Invalid Calibration Vector: acal, Row: 0. Error: 230015035 Cal_db failure See LD for complete error Tue Oct 18 07:20:43 2005 Error: 230015035 Host: ctls3 Process: scanmgr File: cal_db_file_io.c Line: 762 Cal
Applies to
- Models
- HiSpeed, VCT
- Product
- HiSpeed QX/i - GOC2 103L.5_HSQXIM4 103L.4_HSQXIM4 102L.1_HSQXIM3 HiSpeed QX/i - GOC2 Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types GOC2 List of softwares the product can run 103L.5_HSQXIM4 103L.4_HSQXIM4 - obsolete 102L.1_HSQXIM3 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Software
- 103L.5_HSQXIM4 for HiSpeed QXi with Console (GOC2 / Linux Pegasus computer) 103L.5_HSQXIM4 for HiSpeed QXi with Console (GOC2 / Linux Pegasus computer) 103L.5_HSQXIM4 for HiSpeed QXi with Console (GOC2 / Linux Pegasus computer) See FMI 25903 I Helios_H2_O2LESS_CLASSC_Product 09/21/2005 Helios ver 1104L.3 rel 1104L.3_H4.2CLASSC I Helios_CLASSMOC_Product 09/21/2005 CLASSMROOT ver 1204L.3 rel 1204L.3_H4.2CLASSM Class C is 2394897 Order the following P/N: 2401759 (6 pack) Class M is 2394898 Order the following P/N: 2401759 (6 pack) Software Collector 5115431 Recon Patch for GOC2 only 5123440
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.