MR010013 PDS - Troubleshooting and Onsite Recommendations.

Symptom

MR010013 PDS - EQUIP ENV Alert - Chronic Temperature Issues Detected. EQUIP room temperature level is outside of GE specifications at the inlet to the MR Equipment (59-89.6°F* (15-32 °C)*). *Non-condensing humidity with 50% nominal at 65°F (18.3°C). NOTE: The OnWatch Advanced ToolKit detected this problem. This issue is only applicable when detected in this way, and should not be reused in Knowledge objects for normal Reactive service.

Error message

MR010013 PDS - EQUIP ENV Alert - Chronic Temperature Issues Detected. Select this Error only when handling MR010013 PDS Issue Concept. NOTE: The OnWatch Advanced ToolKit detected this problem. This issue is only applicable when detected this way, and should not be reused in Knowledge objects for normal Reactive service.

Applies to

Models
Signa
Product
Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
Software
15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
Area
MR

Full cause & step-by-step resolution